Stellant

Category:

Product Info

What is MEDRAD® Stellant?

MEDRAD® Stellant is a contrast media injection system designed for use during CT examinations, known for its innovative clinical offerings and feature set. It is the injector of choice for advanced CT applications such as cardiac and perfusion imaging. The system is available as a single or a dual head configuration.

How does MEDRAD® Stellant work?

The intuitive touch screen interface allows the user to set the injection protocol needed for the CT examination. MEDRAD® Stellant gives the user the option to store and recall up to 250 CT injection protocols for fast access. Injection protocols are easily adjusted to suit the individual patient by simply selecting the parameter to be changed and entering in the new value. Up to 6 injection phases can be programmed depending upon the area to be imaged.

CT Imaging Solutions start here…

The MEDRAD® Stellant D Dual Syringe CT Injection System is reliable and easy to use. It is an integral piece of the CT imaging solution which also includes the Radimetrics™ Enterprise Platform, Quality Consumables, and Services by Bayer. (Extended warranties including VirtualCareTM Remote Support).

Protocol options for the advanced clinical applications:

  • Designed for complex contrast-enhanced CT protocols
  • Creates a tight contrast bolus
  • Enables precisely timed contrast delivery
  • Clinical benefits of injecting contrast and saline at different ratios to visualize the whole heart
    Clinical research has shown that simultaneous injection of contrast and saline, in user-selected ratios, can:

    • Provide more uniform attenuation of the right and left ventricles
    • Minimize artifacts by achieving proper attenuation levels
    • Visualize the right coronary arteries and right ventricles in a single study by achieving more uniform attenuation

Product Specs

Mechanical Specs

Height Width Depth Weight
Dual Injector Head 15.5 cm (6.1“) 30.7 cm (12.1“) 36.8 cm (14.5“) 8.1 kg (17.9 lb) (without syringe)
Base Unit 29.2 cm (11.5“) 27.9 cm (11.0“) 22.2 cm (8.8“) 6.2 kg (13.6 lb)
Certegra® Workstation (CWS) 34.2 cm (13.5“) 40.0 cm (15.8“) 30.0 cm (10.2“) 8.0 kg (17.6 lb)

 

Injection Specifications

  • Flow Rate (range & increments) 0.1 to 10 mL/sec in 0.1 mL increments
  • Volume (range & increments) 1 mL to syringe capacity in 1 mL increments
  • Programmable Pressure Limit 200 mL syringe: 325 psi, 2241 kPa (psi/kPa) Reminders 0-300 seconds (5 minutes) in 1 second increments
  • Pause 1-900 seconds (15 minutes) in 1 second increments
  • Hold Maximum HOLD time is 20 minutes
  • Syringes (Volume capacity) 200 mL sterile disposable syringe
  • Maximum Number of Phases 6
  • Maximum Number of Protocols 250

 

Miscellaneous Specifications

  • Electrical Requirements (VAC/Hz) 100-240 VAC, 50/60 Hz, 300 Volts-AMPs
  • Syringe Heat Maintainer Range 95 degrees F +/-9 degrees 35 degrees C +/-5 degrees
  • Head Mounting Options Overhead Counterpoise System or Floor Pedestal (Floor Pedestal includes an Integral IV Pole)
  • DualFlow Simultaneous Injection of Contrast and Saline
  • Medrad® XDS Extravasation Extravasation Detector
  • Detection System Extravasation Detection Device
  • Certegra® P3T Cardiac Patient-specific Injection Protocols Application for Cardiac CTA Software
  • Certegra® P3T Abdomen Abdominal Protocol Optimization Software Application
  • Certegra® P3T PA Application Pulmonary Angiography Protocol Optimization Software

Brochure Download