T-MAT G/RA Films
- T-Grain film technology delivers clear and brilliant medical images and minimizes minus-density pre-processing handling artifacts that can compromise radiographic interpretation
- Sensitometrically stable and reliable in automatic processing, and under a variety of processing conditions
- No need for rescreening. Can be used with all green-emitting rare-earth intensifying screens such as LANEX screens or appropriate screens from other manufacturers
- RA (rapid access) technology dramatically reduces processing time and increases productivity
T-Grain Technology:
- Reduced light crossover without a change in film speed provides increased detail
- A standard-speed film that can be used in a high-speed system (up to 600 speed)
- Minimizes minus-density pre-processing handling artifacts that can compromise radiographic interpretation
- Processing Speed: < 40 seconds
- Produced locally, available worldwide (e.g., manufactured in Europe for European use)
Film Storage
- Store unopened boxes in a cool, dry place (50–70°F/10–21°C; 30–50% relative humidity) shielded from x-rays and other penetrating radiation.
Safelighting
- Use safelight filter type GBX-2 or equivalent located at least 4 ft/1.2 m from the film.
Recommended Processing
- T-MAT G/RA film can be processed in less than 40 seconds with a kwik/RA cycle using X-OMAT RA/30 chemicals, or with a standard cycle using RP X-OMAT chemicals.